*0.1% basic accuracy
*100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz test frequencies
*50 m, 100 m, 250 m, 500 m, 1 V rms test levels
*High-speed measurements (25 ms)
*High-speed contact check
*Quick test recovery
*Wide capacitance test range
*Front-end protection
*Built-in comparator
*Optional transformer parameter measurement (HP / Agilent 16060A)
Measurement Ranges:
Parameter Range
|Z|, R, X 1 mohm - 100 Mohms
|Y|, G, B 10 nS - 1000 S
C 1 pF - 1 F
L 10 nH - 100 kH
D 0.0001 - 9.9999
Q 0.1 - 9999.9
θ -180° - +180°
DCR (Opt. 001) 1 mohm - 100 Mohms
N (Opt. 001) 0.9 - 200 (unspecified)
L, M (M is Opt. 001) 1 μH - 100 H
(unspecified)
Δ% -999.99% - 999.99%